• Product description
  • Download files

Rubber dam punch with the function of adjusting the hole diameter.

INTENDED USE
This medical device is intended for use together with rubber dam during dental treatment to isolate the working area. It enables to place rubber dam in oral cavity in right position. Rubber-Dam punch is used for cutting holes in the Rubber-dam. Adjust the hole diameter to fit the tooth the clamp is being fitted on. The punch has 6 holes of different diameter. To isolate anteriors or premolars use the holes of small diameter. Bigger holes should be used to isolate molar teeth. The biggest hole is for the tooth with fixed clamp.

HANDLING DENTAL TOOLS MADE OF STAINLESS STEEL
Special processes have been undertaken during the manufacture of the dental instruments to ensure corrosion-resistance. In spite of this, the durability of these dental instruments depends on their proper use and correct maintenance.

CLEANING, RINSING, WASHING
The packaged instruments are non – sterile. Before sterilisation these new instruments should be washed thoroughly with warm water and using detergent. Before each use these dental instruments must be sterilised.

STERILISATION
All (reusable) dental instruments must be disinfected, then washed out to remove detergent and then dried before sterilization. Ultrasonic cleansing is recommended in particular as the most effective sterilization. Instruments should be sterilized in accordance with national requirements.

Recommended sterilisation methods:

MethodMinimal parameters:
Steam autoclave134°C, for 5 minutes

 

PRECAUTIONS
Adhere to the instructions included and the additional guidelines for cleaning and sterilisation of the instruments.

STORAGE
Dental instruments should be stored in temperatures ranging between 5-30°C.
Before use protect the instrument against contamination or damage of the material structure (cracks, bends, fractures, peeling, etc.).
Failure to adhere to the above stated instructions decreases instrument durability and reduces instrument lifetime.

NON-STERILE – STERILIZE DIRECTLY BEFORE USE

BEFORE USE PLEASE READ THE ATTACHED INSTRUCTIONS CAREFULLY

TO BE USED BY DENTIST ONLY